Abstract

We developed a NIR absorption spectrometry system for detection of toxic substances by using a glass phosphor based LED. Using this NIR absorption spectrometry system, phosphoric acid solution samples were measured by molybdenum-blue method. Absorption band around 900 nm and that around 960 nm were observed. The absorption band around 900 nm increased with increasing of the phosphoric acid concentration. Partial least squares (PLS) analysis was revealed that a lower phosphoric acid concentration limit of 0.01 ppm. Furthermore, Cu dilute solutions were measured. Although there was no clear absorption band related to Cu, PLS analysis was revealed that a lower Cu concentration limit of 0.1 ppm. These results indicated that this NIR absorption spectrometry system is useful for practical applications.

Cite

CITATION STYLE

APA

Uemura, H., Fuchi, S., Kato, R., Amano, K., Hiraizumi, K., Hayase, H., & Takeda, Y. (2015). Development of near-infrared absorption spectrometry system by using NIR wideband glass phosphor LED. In Journal of Physics: Conference Series (Vol. 619). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/619/1/012053