Assessment of the accuracy of an intermittent-scanning continuous glucose monitoring device in patients with type 2 diabetes mellitus undergoing hemodialysis (AIDT2H) study

22Citations
Citations of this article
41Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

FreeStyle Libre has been approved for use in patients undergoing hemodialysis (HD) in Japan, unlike Europe and the United States; however, evidence regarding its accuracy in such patients is sparse. Forty-one participants with type 2 diabetes undergoing HD were recruited. The overall mean absolute relative difference and mean absolute difference were 23.4% and 33.9 mg/dL, respectively. Sensor glucose levels and capillary glucose levels were significantly correlated (r = 0.858, P

Cite

CITATION STYLE

APA

Toyoda, M., Murata, T., Saito, N., Kimura, M., Takahashi, H., Ishida, N., … Hosoda, K. (2021). Assessment of the accuracy of an intermittent-scanning continuous glucose monitoring device in patients with type 2 diabetes mellitus undergoing hemodialysis (AIDT2H) study. Therapeutic Apheresis and Dialysis, 25(5), 586–594. https://doi.org/10.1111/1744-9987.13618