Abstract
FreeStyle Libre has been approved for use in patients undergoing hemodialysis (HD) in Japan, unlike Europe and the United States; however, evidence regarding its accuracy in such patients is sparse. Forty-one participants with type 2 diabetes undergoing HD were recruited. The overall mean absolute relative difference and mean absolute difference were 23.4% and 33.9 mg/dL, respectively. Sensor glucose levels and capillary glucose levels were significantly correlated (r = 0.858, P
Author supplied keywords
Cite
CITATION STYLE
Toyoda, M., Murata, T., Saito, N., Kimura, M., Takahashi, H., Ishida, N., … Hosoda, K. (2021). Assessment of the accuracy of an intermittent-scanning continuous glucose monitoring device in patients with type 2 diabetes mellitus undergoing hemodialysis (AIDT2H) study. Therapeutic Apheresis and Dialysis, 25(5), 586–594. https://doi.org/10.1111/1744-9987.13618