Error introduced by a popular method of processing parallel-disk viscometry data

14Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

The assumptions implicit in the simplified expressions used to convert the torque-rotational speed data of parallel-disk viscometry into rim shear rate and rim shear stress are identified. The rim shear stress generated by the simplified expression is compared against the actual rim shear stress. The error involved is quantified for two standard rheological models and for a set of laboratory data. Under normal operation conditions of parallel-disk viscometers this error was found to be within the acceptable limit. However, for highly shear thinning fluids and for fluids exhibiting yield stress this error can become very large. The suitability of the approximate rim shear stress in wall slip determination is then briefly discussed. © Appl. Rheol.

Cite

CITATION STYLE

APA

Yeow, Y. L., Leong, Y. K., & Khan, A. (2007). Error introduced by a popular method of processing parallel-disk viscometry data. Applied Rheology, 17(6). https://doi.org/10.1515/arh-2007-0022