Performance of an implantable cardioverter-defibrillator lead family

4Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Lead failure is the major limitation in implantable cardioverter-defibrillator (ICD) therapy. Long-term follow-up data for Biotronik Linox ICD leads are limited. Therefore, we analyzed the performance of all these leads implanted at our institution. Materials and Methods: All Linox and Linox Smart ICD leads implanted between 2006 and 2015 were identified. Lead failure was defined as electrical dysfunction (oversensing, abnormal impedance, exit block). Lead survival was described, according to Kaplan–Meier. Associations between lead failure and specific variables were examined. p

Cite

CITATION STYLE

APA

Klampfleitner, S., Mundel, M., Schinke, K., & Neuberger, H. R. (2023). Performance of an implantable cardioverter-defibrillator lead family. Journal of Cardiovascular Electrophysiology, 34(3), 700–709. https://doi.org/10.1111/jce.15804

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free