Abstract
Background: Lead failure is the major limitation in implantable cardioverter-defibrillator (ICD) therapy. Long-term follow-up data for Biotronik Linox ICD leads are limited. Therefore, we analyzed the performance of all these leads implanted at our institution. Materials and Methods: All Linox and Linox Smart ICD leads implanted between 2006 and 2015 were identified. Lead failure was defined as electrical dysfunction (oversensing, abnormal impedance, exit block). Lead survival was described, according to Kaplan–Meier. Associations between lead failure and specific variables were examined. p
Author supplied keywords
Cite
CITATION STYLE
Klampfleitner, S., Mundel, M., Schinke, K., & Neuberger, H. R. (2023). Performance of an implantable cardioverter-defibrillator lead family. Journal of Cardiovascular Electrophysiology, 34(3), 700–709. https://doi.org/10.1111/jce.15804