Abstract
Fibreoptic intubation remains a key technique for the management of difficult intubation. We randomly compared the second generation single-use Ambu® aScope™ 2 videoscope with a standard re-usable flexible intubating fibrescope in 50 tracheal intubations in patients with a difficult airway simulated by a semirigid collar. All patients' tracheas were intubated successfully with the aScope 2 or the re-usable fibrescope. The median (IQR [range]) time to intubate was significantly longer with the aScope 2 70 (55-97 [41?-226]) s vs 50 (40-59 [27-175]) s, p = 0.0003) due to an increased time to see the carina. Quality of vision was significantly lower with the aScope 2 (excellent 24 (48%) vs 49 (98%), p = 0.0001; good 22 (44%) vs 1 (2%), p = 0.0001; poor 4 (8%) vs 0, p = 0.12) but with no difference in the subjective ease to intubate (easy score of 31 (62%) vs 38 (76%), p = 0.19; intermediate 12 (24%) vs 7 (14%), p = 0.31; difficult 7 (14%) vs 5 (5%), p = 0.76). The longer times to intubate and the poorer scores for quality of vision do not support the use of the single-use aScope 2 videoscope as an alternative to the re-usable fibrescope. © Anaesthesia © 2012 The Association of Anaesthetists of Great Britain and Ireland.
Cite
CITATION STYLE
Krugel, V., Bathory, I., Frascarolo, P., & Schoettker, P. (2013). Comparison of the single-use Ambu® aScopeTM 2 vs the conventional fibrescope for tracheal intubation in patients with cervical spine immobilisation by a semirigid collar. Anaesthesia, 68(1), 21–26. https://doi.org/10.1111/anae.12044