Study of a spherical phantom for Gamma knife dosimetry

12Citations
Citations of this article
22Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Four 16 cm diameter spherical phantoms were modeled in this study: A homogenous water phantom, and three water phantoms with 1 cm thick shell each made of different materials (PMMA, Plastic WaterTM and polystyrene). The PENELOPE Monte Carlo code was utilized to simulate photon beams from the Leksell Gamma Knife (LGK) unit and to determine absorbed dose to water (Dw) from a single 18 mm beam delivered to each phantom. A score spherical volume of 0.007 cm3 was used to simulate the dimensions of the sensitive volume of the Exradin A-16 ionization chamber, in the center of the phantom. In conclusion, the PMMA shell filled with water required a small correction for the determination of the absorbed dose, while remaining within the statistical uncertainty of the calculations (± 0.71). Plastic WaterTM and polystyrene shells can be used without correction. There is a potential advantage to measuring the 4 mm helmet output using these spherical water phantoms.

Cite

CITATION STYLE

APA

Zhu, D., Austerlitz, C., Benhabib, S., Mota, H., Allison, R. R., & Campos, D. (2010). Study of a spherical phantom for Gamma knife dosimetry. Journal of Applied Clinical Medical Physics, 11(2), 222–229. https://doi.org/10.1120/jacmp.v11i2.3130