Measurement of a PAGAT gel dosimeter by ultrasound computed tomography

5Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

In this work we used a 3D quantitative CT ultrasound imaging system to characterise polymer gel dosimeters. The system comprised of two identical 5 MHz 128 element phased-array ultrasound transducers co-axially aligned and submerged in water as a coupling agent. Rotational and translational movement of the gel dosimeter sample between the transducers were performed using a robotic arm. Ultrasound signals were generated and received using an Olympus Omniscan unit. Dose sensitivity of attenuation and time of flight ultrasonic parameters were assessed using this system.

Cite

CITATION STYLE

APA

Khoei, S., Trapp, J. V., & Langton, C. M. (2013). Measurement of a PAGAT gel dosimeter by ultrasound computed tomography. In Journal of Physics: Conference Series (Vol. 444). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/444/1/012083

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free