Abstract
Ibuprofen membrane electrodes based on different plasticizers: diisobutyl phthalate (DIBP), o-nitrophenyloctyl ether (o-NPOE), dioctyl sebacate (DOS) and tetraoctylammonium 2-(4-isobutylphenyl)propionate were prepared. All electrodes show: a near Nernstian slope of characteristic (58.3-60.9 mV decade-1) in the measurement range (10-4-10-1 mol L-1), limit of detection (5.0×10-5 mol L-1), really long lifetime (12 months), dependence of the electrode potential on pH (5.5-9.0), reproducibility of potential (0.6-1.2 mV) and selectivity coefficients in relation to some organic and inorganic anions. The electrodes were applied for the determination of ibuprofen in tablets by the calibration curve method and the standard addition method. © Versita Warsaw and Springer-Verlag Wien 2010.
Author supplied keywords
Cite
CITATION STYLE
Lenik, J., & Wardak, C. (2010). Properties of ibuprofen ion-selective electrodes based on the ion pair complex of tetraoctylammonium cation. Central European Journal of Chemistry, 8(2), 382–391. https://doi.org/10.2478/s11532-010-0005-3