Abstract
Purpose: To describe a new time-saving threshold visual field–testing strategy—Swedish Interactive Thresholding Algorithm (SITA) Faster, which is intended to replace SITA Fast—and to report on a clinical evaluation of this new strategy. Design: Description and validity analysis for modifications applied to SITA Fast. Methods: Five centers tested 1 eye of each of 126 glaucoma and glaucoma suspect patients with SITA Faster, SITA Fast, and SITA Standard at each of 2 visits. Outcomes included test time, mean deviation, and the visual field index (VFI), significant test points in probability maps, and intertest threshold variability. Results: Mean (standard deviation) test times were 171.9 (45.3) seconds for SITA Faster, 247.0 (56.7) for SITA Fast, and 369.5 (64.5) for SITA Standard (P
Cite
CITATION STYLE
Heijl, A., Patella, V. M., Chong, L. X., Iwase, A., Leung, C. K., Tuulonen, A., … Bengtsson, B. (2019). A New SITA Perimetric Threshold Testing Algorithm: Construction and a Multicenter Clinical Study. American Journal of Ophthalmology, 198, 154–165. https://doi.org/10.1016/j.ajo.2018.10.010