Elimination or memory effects and simultaneous determination of B, Br, I, Hg, Bi, and Sb in groundwater samples by inductively coupled plasma mass spectrometry

8Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

Abstract

In this study, elimination of the memory effects of B, Br, I, Hg, Bi, and Sb and their simultaneous determination in groundwater sampls were studied in detail by inductively coupled plasma mass spectrometry (ICP-MS). The memory effects were eliminated with a wash run of 4.5 minutes with a 0.5% NH 3-H2O/0.5% NaF/10 mg I-1 Au solution. Good agreement between the measured and the reference values of the CRM samples indicated the effectiveness of this method. A good linear calibration curve was obtained with a correlation coefficient of 0.9999. Therefore, a method for the simultaneous determination of B, Br, I, Hg, Sb, and Bi in groundwater samples by ICP-MS was established.

Cite

CITATION STYLE

APA

Zhang, J. Y., Li, G. M., & Guo, W. (2014). Elimination or memory effects and simultaneous determination of B, Br, I, Hg, Bi, and Sb in groundwater samples by inductively coupled plasma mass spectrometry. Atomic Spectroscopy, 35(3), 97–102. https://doi.org/10.46770/as.2014.03.001