Evaluation of shock tube retrofitted with fast‐opening valve for dynamic pressure calibration

21Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

Abstract

Accurate dynamic pressure measurements are increasingly important. While traceability is lacking, several National Metrology Institutes (NMIs) and calibration laboratories are currently establishing calibration capacities. Shock tubes generating pressure steps with rise times below 1 μs are highly suitable as standards for dynamic pressures in gas. In this work, we present the results from applying a fast‐opening valve (FOV) to a shock tube designed for dynamic pressure measurements. We compare the performance of the shock tube when operated with conventional single and double diaphragms and when operated using an FOV. Different aspects are addressed: shock‐wave formation, repeatability in amplitude of the realized pressure steps, the assessment of the required driver pressure for realizing nominal pressure steps, and economy. The results show that using the FOV has many advantages compared to the diaphragm: better repeatability, eight times faster to operate, and enables automation of the test sequences.

Cite

CITATION STYLE

APA

Amer, E., Wozniak, M., Jönsson, G., & Arrhén, F. (2021). Evaluation of shock tube retrofitted with fast‐opening valve for dynamic pressure calibration. Sensors, 21(13). https://doi.org/10.3390/s21134470

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free