Dose integration and dose rate characteristics of a NiPAM polymer gel MRI dosimeter system

18Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The normoxic polymer gel dosimeter based on N-isopropyl acrylamide (NiPAM) is a promising full 3D-dosimeter with high spatial resolution and near tissue equivalency. NiPAM gel samples were irradiated to different doses using a linear accelerator. The absorbed dose was evaluated using MRI and statistical significance of the analysed data was calculated. The analysis was carried out using an in-house developed software. It was found that the gel dosimeter responded linearly to the absorbed dose. The gel exhibited a dose rate dependence, as well as a dependence on the sequential beam irradiation scheme. A higher dose rate, as well as a higher dose per sequential beam, resulted in a lower dose response.