Objectives: Handheld ultrasound devices (HUDs) have previously been limited to grayscale imaging without options for left ventricle (LV) quantification. We aimed to study the feasibility and reliability of automatic measurements of mitral annular plane systolic excursion (MAPSE) by HUDs. Methods: An algorithm that automatically measured MAPSE from live grayscale recordings was implemented in a HUD. Twenty patients at a university hospital were examined by either a cardiologist or a sonographer. Standard echocardiography using a high-end scanner was performed. The apical 4-chamber view was recorded 4 times by both echocardiography and the HUD. MAPSE was measured by M-mode and color tissue Doppler (cTD) during echocardiography and automatically by the HUD. Results: The automatic method underestimated mean MAPSE ± SD versus M-mode (9.6 ± 2.2 versus 10.9 ± 2.6 mm; difference, 1.2 ± 1.4 mm, P
CITATION STYLE
Magelssen, M. I., Palmer, C. L., Hjorth-Hansen, A., Nilsen, H. O., Kiss, G., Torp, H., … Dalen, H. (2021). Feasibility and Reliability of Automatic Quantitative Analyses of Mitral Annular Plane Systolic Excursion by Handheld Ultrasound Devices: A Pilot Study. Journal of Ultrasound in Medicine, 40(2), 341–350. https://doi.org/10.1002/jum.15408