Modular titanium alloy neck failure in total hip replacement: Analysis of a relapse case

13Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.

Abstract

Modular neck hip prosthesis born in the 1990 with the aim of allowing the surgeon to modify CCD angle, offset and femoral anteversion intra-operatively restoring patient's original biomechanics. In order to achieve the best biomechanics of the reconstructed hip, preoperative planning is essential. In the last few years modularity has been questioned and an argument made for the return to mono block stems due to events of breakage or disconnection of modular components. Fretting or crevice corrosion may lead to failure of such modular device due to the contamination inside the modular coupling or to high loads. We present a case of repetitive modular femoral neck prosthesis fracture.

Cite

CITATION STYLE

APA

Ceretti, M., & Falez, F. (2016). Modular titanium alloy neck failure in total hip replacement: Analysis of a relapse case. SICOT-J, 2. https://doi.org/10.1051/sicotj/2016009