Abstract
The use of subcutaneous implantable cardioverter-defibrillators is a novel option for preventing arrhythmia-mediated cardiac death in patients who are at risk of endovasculardevice infection or in whom venous access is difficult. However, the potential for electromagnetic interference between subcutaneous defibrillators and left ventricular assist devices is largely unknown. We report the case of a 24-year-old man in whom we observed no electromagnetic interference between a subcutaneous implanted cardioverterdefibrillator and a HeartMate II Left Ventricular Assist System, at 3 different pump speeds. To our knowledge, this is the first report of such findings in this circumstance.
Author supplied keywords
Cite
CITATION STYLE
Raman, A. S., Shabari, F. R., Kar, B., Loyalka, P., & Hariharan, R. (2016). No electromagnetic interference occurred: In a patient with a heartmate ii left ventricular assist system and a subcutaneous implantable cardioverter-defibrillator. Texas Heart Institute Journal, 43(2), 183–185. https://doi.org/10.14503/THIJ-14-4795