Abstract
SwissFEL requires the monitoring of the photon spectral distribution at a repetition rate of 100 Hz for machine optimization and experiment online diagnostics. The Photon Single Shot Spectrometer has been designed for the photon energy range of 4 keV to 12 keV provided by the Aramis beamline. It is capable of measuring the spectrum in a non-destructive manner, with an energy resolution of Δ E/E = (2-5) × 10-5 over a bandwidth of 0.5% on a shot-to-shot basis. This article gives a detailed description about the technical challenges, structures, and considerations when building such a device, and to further enhance the performance of the spectrometer.
Author supplied keywords
Cite
CITATION STYLE
Rehanek, J., Makita, M., Wiegand, P., Heimgartner, P., Pradervand, C., Seniutinas, G., … Juranić, P. (2017). The hard X-ray Photon Single-Shot Spectrometer of SwissFEL - Initial characterization. Journal of Instrumentation, 12(5). https://doi.org/10.1088/1748-0221/12/05/P05024