Peak flow measurements in patients with severe aortic stenosis: a prospective comparative study between cardiovascular magnetic resonance 2D and 4D flow and transthoracic echocardiography

11Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Aortic valve stenosis (AS) is the most prevalent valvular disease in the developed countries. Four-dimensional (4D) flow cardiovascular magnetic resonance (CMR) is an emerging imaging technique, which has been suggested to improve the evaluation of AS severity compared to two-dimensional (2D) flow and transthoracic echocardiography (TTE). We investigated the reliability of CMR 2D flow and 4D flow techniques in measuring aortic transvalvular peak systolic flow in patients with severe AS. Methods: We prospectively recruited 90 patients referred for aortic valve replacement due to severe AS (73.3 ± 11.3 years, aortic valve area 0.7 ± 0.1 cm2, and 54/36 tricuspid/bicuspid), and 10 non-valvular disease controls. All the patients underwent echocardiography and 2D flow and 4D flow CMR. Peak flow velocity measurements were compared using Wilcoxon signed rank sum test and Bland–Altman analysis. Results: 4D flow underestimated peak flow velocity in the AS group when compared with TTE (bias − 1.1 m/s, limits of agreement ± 1.4 m/s) and 2D flow (bias − 1.2 m/s, limits of agreement ± 1.6 m/s). The differences between values obtained by TTE (median 4.3 m/s, range 2.7–6.1 m/s) and 2D flow (median 4.5 m/s, range 2.9–6.5 m/s) compared to 4D flow (median 3.1 m/s, range 1.7–5.1 m/s) were significant (p < 0.001). The difference between 2D flow and TTE were insignificant (bias 0.07 m/s, limits of agreement ± 1.5 m/s). In non-valvular disease controls, peak flow velocity was measured higher by 4D flow than 2D flow (1.4 m/s, 1.1–1.7 m/s and 1.3 m/s, 1.1–1.5 m/s, respectively; bias 0.2 m/s, limits of agreement ± 0.16 m/s). Conclusions: CMR 4D flow significantly underestimates systolic peak flow velocity in patients with severe AS. 2D flow, in turn, estimated the AS velocity accurately, with measured peak flow velocities comparable to TTE.

Cite

CITATION STYLE

APA

Hälvä, R., Vaara, S. M., Peltonen, J. I., Kaasalainen, T. T., Holmström, M., Lommi, J., … Syväranta, S. (2021). Peak flow measurements in patients with severe aortic stenosis: a prospective comparative study between cardiovascular magnetic resonance 2D and 4D flow and transthoracic echocardiography. Journal of Cardiovascular Magnetic Resonance, 23(1). https://doi.org/10.1186/s12968-021-00825-1

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free