Abstract
An accurate, sensitive and reproducible high performance liquid chromatographic (HPLC) method for the quantitation of doxycycline in turkey serum has beendeveloped and validated. The doxycycline standard was eluted from a C18 column (4.6 mm i.d., 250 mm, 5 µm particle size) at 15 ⁰C with a mobile phase consisting of 5% acetic acid : methanol : acetonitrile (55:20:25, v/v/v). A UV detector set at 347 nm was used to monitor the effluent. Retention time of doxycycline was approximately at 4 min and the run time was 10 min. The column temperature adjusted at 15 ⁰C with a flow rate 1ml/min and injection volume 25 µl. Doxycycline was extracted from serum by ACN. Linearity was confirmed in the concentration range 0.1-2 µg/ml, had regression coefficients (R2) 0.9999 for doxycycline quantification in turkey serum. The analytical conditions gave a recoveries in a range of 97.57% to 99.83%. The limit of detection (LOD) was 0.035 µg/ mL and limit of quantification (LOQ) was 0.108 µg/ml. The relative standard deviation (RSD %) was less than 2% in all cases. The developed method was accurate, precise and rapid for simultaneous estimation of doxycycline in turkey serum samples.