Development and long-term evaluation of a new 68Ge/68Ga generator based on nano-SnO2 for PET imaging

24Citations
Citations of this article
66Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Radionuclide generator systems can routinely provide radionuclides on demand such as 68Ga produced by a 68Ge/68Ga generator without the availability of an on-site accelerator or a research reactor. Thus, in this work nano-SnO2 was used to develop a new 68Ge/68Ga generator which was evaluated over a period of 17 months and 305 elution cycles. The elution yield was 91.1 ± 1.8% in the first 7 mL (1 M HCl as eluent) when the generator was new and then it decreased with time and use to 73.8 ± 1.9%. Around 80% of the elutable 68Ga activity was obtained in 1 mL and the 68Ge content in the eluate did not exceed 1 × 10–4% over the investigation period when it was eluted regularly. The described generator provided adequate results for radiolabelling of DOTA-TOC with direct use of eluate. In addition, [68Ga]Ga-DOTA-TOC was tested satisfactorily for in vivo tumor detection by microPET/CT imaging in a lung cancer mouse model.

Cite

CITATION STYLE

APA

Romero, E., Martínez, A., Oteo, M., Ibañez, M., Santos, M., & Morcillo, M. Á. (2020). Development and long-term evaluation of a new 68Ge/68Ga generator based on nano-SnO2 for PET imaging. Scientific Reports, 10(1). https://doi.org/10.1038/s41598-020-69659-8