Hydroxyapatite-coated versus grit-blasted femoral stems

46Citations
Citations of this article
20Readers
Mendeley users who have this article in their library.

Abstract

We have carried out a prospective, randomised study designed to compare the long-term stability of the stem of cementless femoral implants with differing surface configurations. A total of 50 hips (46 patients) was randomised into two groups, according to whether the medullary stem had been grit blasted (GB) or coated with hydroxyapatite (HA). Both femoral prostheses were of the same geometrical design. We used Ein Bild Roentgen Analyse femoral component analysis (EBRA-FCA) to assess the stability of the stem. The mean follow-up was for 8.66 years. The mean migration of the stem was 1.26 mm in the HA group compared with 2.57 mm in the GB group (Mann-Whitney U test, p = 0.04). A mixed model ANOVA showed that the development of subsidence was statistically different in the two groups during the first 24 months. After this subsidence increased in both groups with no difference between them. Our results indicate that, with the same design of stem, HA coating enhanced the stability of the femoral stem when compared with GB stems.

Cite

CITATION STYLE

APA

Hamadouche, M., Witvoet, J., Porcher, R., Meunier, A., Sedel, L., & Nizard, R. (2001). Hydroxyapatite-coated versus grit-blasted femoral stems. Journal of Bone and Joint Surgery - Series B, 83(7), 979–987. https://doi.org/10.1302/0301-620X.83B7.11478

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free