This study compared two bedside methods recommended for the detection of low concentrations of heparin and the activated partial thromboplastin time (APTT), with reference to a laboratory measure of heparin concentration. Patients undergoing cardiopulmonary bypass had blood drawn at four stages when low levels of heparin could be expected. At each stage four tests were performed: whole blood clotting time using a Hemochron analyser with a Saline-Rinsed test cartridge, whole blood clotting time using a Hemotec analyser with a High Range Heparinase test cartridge, APTT, and heparin concentration bypolybrene neutralization. Thirty patients were studied. The sensitivity of the Saline-Rinsed Hemochron, Hemotec High Range Heparinase, and APTT in detecting concentrations of heparin less than 1 U/ml was 38%, 40% and 97%, respectively, while specificities were 87%, 90%, and 30%, respectively. Neither the Saline Rinsed Hemochron, nor the Hemotec Heparinase cartridge reliably detected concentrations of heparin less than 1 U/ml.
CITATION STYLE
Weightman, W. M., Clapin, A. O. G., Michalopoulis, N., & Gibbs, N. M. (1996). Sensitivity and Specificity of the Hemochron, Hemotec and APTT for the Detection of Low Concentrations of Heparin. Anaesthesia and Intensive Care, 24(4), 472–476. https://doi.org/10.1177/0310057x9602400410