Objective: The objectives of this study were to evaluate the accuracy of computed tomography (CT)-based navigation and to investigate whether the level of surgeon experience affects the accuracy of cup positioning under navigation. Methods: This study investigated 117 hips in 103 patients who underwent primary total hip arthroplasty (THA) by 7 surgeons using a CT-based navigation system. Pre- and postoperative CT images were matched using a volume registration technique. Postoperative cup angles and positions were then measured using the same pelvic coordinates, and results were compared for experienced and inexperienced surgeons. Results: The mean absolute error of the cup angle was 1.8 ± 1.6°for inclination and 1.2 ± 1.1°for anteversion. The mean absolute errors of cup position were 1.9 ± 1.5 mm, 1.4 ± 1.2 mm, and 1.9 ± 1.3 mm on the x-, y- and z-axes, respectively. No significant differences in accuracy were identified between experienced and inexperienced surgeons. Conclusions: The absolute spatial error of cup position was ≤2 mm for each axis, and the angle error was ≤2°for the angles of inclination and anteversion. This navigation system could therefore help surgeons perform accurate cup placement irrespective of the surgeon's level of experience. © 2013 Informa UK Ltd.
CITATION STYLE
Iwana, D., Nakamura, N., Miki, H., Kitada, M., Hananouchi, T., & Sugano, N. (2013). Accuracy of angle and position of the cup using computed tomography-based navigation systems in total hip arthroplasty. Computer Aided Surgery, 18(5–6), 187–194. https://doi.org/10.3109/10929088.2013.818713