Micro-dislodgement during transcatheter aortic valve implantation with a contemporary self-expandable prosthesis

8Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

Abstract

Objectives To evaluate the incidence, risk factors and the clinical outcome of micro-dislodgement (MD) with a contemporary self-expandable prosthesis during transcatheter aortic valve implantation. Methods MD was defined as movement of the prosthesis of at least 1.5 mm upwards or downwards from its position directly before release compared to its final position. Patients were grouped according to the occurrence (+MD) or absence (-MD) of MD. Baseline characteristics, imaging data and outcome parameters were retrospectively analyzed. Results We identified 258 eligible patients. MD occurred in 31.8% (n = 82) of cases with a mean magnitude of 2.8 mm ± 2.2 in relation to the left coronary cusp and 3.0 mm ± 2.1 to the non-coronary cusp. Clinical and hemodynamic outcomes were similar in both groups with consistency over a follow-up period of three months. A larger aortic valve area (AVA) (-MD vs. +MD: 0.6 cm2 ± 0.3 vs. 0.7cm2 ± 0.2; p = 0.014), was the only independent risk factor for the occurrence of MD in a multivariate regression analysis (OR 5.3; 95% CI: 1.1–24.9; p = 0.036). Conclusions MD occurred in nearly one third of patients and did not affect clinical and hemodynamic outcome. A larger AVA seems to be a potential risk factor for MD.

Cite

CITATION STYLE

APA

Hellhammer, K., Piayda, K., Afzal, S., Veulemans, V., Hennig, I., Makosch, M., … Zeus, T. (2019). Micro-dislodgement during transcatheter aortic valve implantation with a contemporary self-expandable prosthesis. PLoS ONE, 14(11). https://doi.org/10.1371/journal.pone.0224815

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free