Abstract
This work presents the development and characterization of a flexible capacitive electrode for non-contact ECG acquisition, fabricated using a simple and cost-effective method from readily available materials. The electrode consists of a multilayer structure with a copper conductor laminated by a polyimide (Kapton®) dielectric layer on a polyurethane support. The impedance and capacitance of the electrode were evaluated under varying textile moisture levels with artificial sweat, as well as after exposure to common disinfectants including ethyl alcohol and iodine tincture. Electrochemical impedance spectroscopy (EIS) and broadband impedance measurements (10−1–105 Hz) confirmed stable capacitive behavior, moderate sensitivity to moisture, and chemical stability of the Kapton–copper interface under conditions simulating repeated use. A custom front-end readout circuit was implemented to demonstrate through-textile ECG signal acquisition. Simulator tests reproduced characteristic waveform patterns, and preliminary volunteer recordings confirmed the feasibility of through-textile acquisition. These results highlight the promise of the electrode as a low-cost platform for future wearable biosignal monitoring technical research.