Fast determination of minoxidil by photometric flow titration

17Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

A photometric flow titration based on the redox reaction between KMnO 4 and minoxidil is described. The best titration results were observed at 3.20 × 10-4 mol L-1 KMnO4 and 1.00 × 10-3 mol L-1 minoxidil, using the minoxidil solutions as titrant. The flow rate was fixed at 17 mL min-1 and the titrant was added to the system in aliquots of 500 μL, the color changes were monitored at 550 nm. The method was applied to commercial samples and compared with the results from a chromatographic procedure. Recoveries from 97.6 to 102.8 % were observed depending on the sample. Comparison with the chromatographic procedure reveled relative errors of 3.5 - 4.0%.

Cite

CITATION STYLE

APA

De Sousa, R. A., Semaan, F. S., Baio, J. A. F., & Cavalheiro, E. T. G. (2005). Fast determination of minoxidil by photometric flow titration. Ecletica Quimica, 30(3), 79–84. https://doi.org/10.1590/s0100-46702005000300010