Abstract
A novel sequential-injection system was developed for determination of iodide at very low concentrations by using a kinetic method. The method is based on the catalytic effect of iodide on the redox reaction between Ce4+ and As3+ first described by Sandell and Kolthoff. The calibration curve is constructed by measuring the decrease of Ce4+ absorbance versus iodide concentration with a delay time of 30 s. The detection limit is 1.5 μg/L, the working temperature is 45°C, and the working range is 0.002-0.5 mg/L. Reasonable agreement was obtained when the method was applied to pharmaceutical and drinking water samples. The method has a sample throughput of approximately 15/h.
Cite
CITATION STYLE
Erustes, J. A., Forteza, R., & Cerdà, V. (2001). Sequential-injection procedure for determination of iodide in pharmaceutical and drinking water samples by catalytic reaction with spectrophotometric detection. Journal of AOAC International, 84(2), 337–341. https://doi.org/10.1093/jaoac/84.2.337