Abstract
A total of 7,190 blond culture sets were obtained from adult patients with a suspected bloodstream infection. A 20-ml sample of blood was distributed equally between the aerobic FAN bottle which was monitored in the BacT/Alert system and a Plus Aerobic/F bottle which was monitored in the BACTEC 9240 system. A total of 988 positive cultures were obtained from 483 patients; however, only 453 positive cultures from 173 patients met the criteria for volume (≥8 ml per bottle) and clinical significance on the basis of concurrent case review required for data analysis. There were 25 and 68 false positives from the FAN and Plus Aerobic/F bottles, respectively. There were no statistically significant differences between systems in the number of positive cultures or septic episodes by species; however, the total number of Enterobacteriaceae and Pseudomonas aeruginosa isolates combined was significantly greater in the FAN bottle (P = 0.04). Detection times did not differ significantly between systems for positive cultures; however, episodes of Staphylococcus aureus bacteremia were detected significantly more rapidly from the FAN bottle (P = 0.005). There was no significant difference between systems in the detection of bloodstream infections in patients receiving antibiotics at the time of blood culture.
Cite
CITATION STYLE
Pohlman, J. K., Kirkley, B. A., Easley, K. A., Basille, B. A., & Washington, J. A. (1995). Controlled clinical evaluation of BACTEC Plus Aerobic/F and BacT/Alert aerobic FAN bottles for detection of bloodstream infections. Journal of Clinical Microbiology, 33(11), 2856–2858. https://doi.org/10.1128/jcm.33.11.2856-2858.1995