Bending Angle Sensor Based on Double-Layer Capacitance Suitable for Human Joint

23Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Goal: To develop bending angle sensors based on double-layer capacitance for monitoring joint angles during cycling exercises. Methods: We develop a bending angle sensor based on double-layer capacitive and conducted three stretching, bending, and cycling tests to evaluate its validity. Results: We demonstrate that the bending angle sensor based on double-layer capacitance minimizes the change in the capacitance difference in the stretching test. The hysteresis and root mean square error (RMSE) compared with the optical motion capture show hysteresis: 8.0% RMSE and 3.1° in the bending test. Moreover, a cycling experiment for human joint angle measurements confirm the changes in accuracy. The RMSEs ranged from 4.7° to 7.0°, even when a human wears leggings fixed with the developed bending-angle sensor in the cycling test. Conclusion: The developed bending angle sensor provides a practical application of the quantitative and observational evaluation tool for knee joint angles.

Cite

CITATION STYLE

APA

Goto, D., Sakaue, Y., Kobayashi, T., Kawamura, K., Okada, S., & Shiozawa, N. (2023). Bending Angle Sensor Based on Double-Layer Capacitance Suitable for Human Joint. IEEE Open Journal of Engineering in Medicine and Biology, 4, 129–140. https://doi.org/10.1109/OJEMB.2023.3289318