Experiences of a medium-sized uhv system for pressures below 10-11 mbar

8Citations
Citations of this article
1Readers
Mendeley users who have this article in their library.
Get full text

Abstract

There are several modes of operation planned for the CRYRING heavy ion accelerator/storage system. From vacuum considerations, it may be reasonable to separate these modes into: (i) acceleration; (ii) storage at high energies; (iii) storage at low energies. The vacua required may be calculated from simple considerations using interpolated electron capture cross-sections. From this starting point, a test system was designed and constructed. This is the topic of the present paper together with some initial results showing that it is possible to reach a vacuum in the lower 10-12 torr region. © 1987.

Cite

CITATION STYLE

APA

Lindblad, T., Bagge, L., Bjon, J., & Levén, S. (1987). Experiences of a medium-sized uhv system for pressures below 10-11 mbar. Vacuum, 37(3–4), 293–296. https://doi.org/10.1016/0042-207X(87)90012-1