Abstract
Purpose: To establish the feasibility of intracranial time-of-flight (TOF) MR angiography (MRA) at 7T using phased array coils and to compare its performance to 3T. Materials and Methods: In an initial study, rive normal volunteers were scanned at 7T and 3T using eight-channel coils and standard acquisition parameters from a clinical TOF protocol. In a second study three additional studies were performed at 7T and 3T using empirically optimized 7T parameters. Contrast-to-noise (CNR) values were measured for major vessel segments. Results: All measurements documented CNR increases at 7T, with a mean increase of 83% in the initial study and 88% in the second study. The CNR values achieved using the latter protocol were similar to the values obtained in the initial study, despite the 42% reduction expected due to the higher spatial resolution. CNR in the smaller peripheral vessels was increased dramatically, resulting in excellent visualization at high resolution. Conclusion: TOF MRA at 7T demonstrated improved visualization of the intracranial vasculature, particularly the smaller peripheral vessels, and may benefit studies of small aneurysms, atherosclerosis, vasospasm, and vasculitis. © 2007 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Von Morze, C., Xu, D., Purcell, D. D., Hess, C. P., Mukherjee, P., Saloner, D., … Vigneron, D. B. (2007). Intracranial time-of-flight MR angiography at 7T with comparison to 3T. Journal of Magnetic Resonance Imaging, 26(4), 900–904. https://doi.org/10.1002/jmri.21097