Computational design and evaluation of a quad-MOSFET device for quality control of therapeutic accelerator-based neutron beams

3Citations
Citations of this article
3Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Accurate real-time monitoring of neutron beams and distinguishing between thermal, epithermal and fast neutron components in the presence of a photon background is crucial for the effectiveness of accelerator-based boron neutron capture therapy (AB-BNCT). In this work, we propose an innovative quadruple metal–oxide–semiconductor field-effect transistor (MOSFET) device for real-time, cost-effective beam quality control; one detector is kept uncovered while the other three are covered with either a B4C, cadmium and B4C or polyethylene converter. Individual MOSFET converter configurations were optimised via Monte Carlo simulations to maximise signal selectivity across neutron energy spectra. Results demonstrate the quad-MOSFET device's efficacy in quantifying changes in neutron flux, underscoring its potential as a useful instrument in the AB-BNCT quality control process.

Cite

CITATION STYLE

APA

Jakubowski, K., Vohradsky, J., Chacon, A., Franklin, D. R., Tran, L. T., Guatelli, S., … Rosenfeld, A. (2024). Computational design and evaluation of a quad-MOSFET device for quality control of therapeutic accelerator-based neutron beams. Radiation Measurements, 177. https://doi.org/10.1016/j.radmeas.2024.107253