The aim of this study is to test the cleaning effect and surface modification of a new implant surface treatment on explanted dental implants and titanium discs. It is a modified air powder abrasive (APA) treatment applied using osteoconductive powders. Twenty-eight in vitro Ca-precipitated organic film-coated titanium discs and 13 explanted dental implants were treated. In a 2-step approach, 3 powders were used: hydroxylapatite (HA) and biomimetic calcium phosphate (BioCaP), which are osteoconductive, and erythritol, which is not. APA treatment was applied. (Air pressure: 2.4 bar; water flow for cleaning: 41.5 ml/min, for Coating 1: 2.1 ml/min, and for Coating 2: 15.2 ml/min.) The test groups were as follows: Group 1: HA cleaning + BioCaP Coating 1; Group 2: HA cleaning + BioCaP Coating 2; Group 3: erythritol cleaning + BioCaP Coating 1; Group 4: erythritol cleaning + BioCaP Coating 2; Group 5: HA cleaning; Group 6: erythritol cleaning; and control: no powder. Cleaned areas were calculated by point counting method. Surface changes and chemical content were evaluated using light microscopy, scanning electron microscopy, and energy-dispersive X-ray spectroscopy. Cleaning effect between groups was compared by a pairwise Student's t test. The significance level was fixed at p
CITATION STYLE
Tastepe, C. S., Lin, X., Werner, A., Donnet, M., Wismeijer, D., & Liu, Y. (2018). Cleaning effect of osteoconductive powder abrasive treatment on explanted human implants and biofilm-coated titanium discs. Clinical and Experimental Dental Research, 4(1), 25–34. https://doi.org/10.1002/cre2.100