Abstract
A simple and sensitive method was developed for fexofenadine determination in human plasma by liquid chromatography with ultraviolet detection. Satisfactory separation was achieved on a Hypersil® BDS C18 column (250 × 4.6 mm, 5μm) using a mobile phase comprising 20 mm sodium dihydrogen phosphate-2 hydrate (pH adjusted to 3 with phosphoric acid)-acetonitrile at a ratio of 52:48, v/v. The elution was isocratic at ambient temperature with a flow rate of 1.0 mL/min. The UV detector was set at 215 nm for the drug and 330 nm for the internal standared (tinidazole). The total time for a chromatographic separation was ~6.5 min. Linearity was demonstrated over the concentration range 0.01-4 μg/mL. The observed within- and between-day assay precision ranged from 0.346 to 13.6%; accuracy varied between 100.4 and 111.2%. This method was successfully applied for therapeutic drug monitoring in patients treated with clinical doses of fexofenadine and for pharmacokinetic studies.