Mechanism of microwave sterilization in the dry state.

106Citations
Citations of this article
79Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

With an automated computerized temperature control and a specialized temperature measurement system, dry spores of Bacillus subtilis subsp. niger were treated with heat simultaneously in a convection dry-heat oven and a microwave oven. The temperature of the microwave oven was monitored such that the temperature profiles of the spore samples in both heat sources were nearly identical. Under these experimental conditions, we unequivocally demonstrated that the mechanism of sporicidal action of the microwaves was caused solely by thermal effects. Nonthermal effects were not significant in a dry microwave sterilization process. Both heating systems showed that a dwelling time of more than 45 min was required to sterilize 10(5) inoculated spores in dry glass vials at 137 degrees C. The D values of both heating systems were 88, 14, and 7 min at 117, 130, and 137 degrees C, respectively. The Z value was estimated to be 18 degrees C.

Cite

CITATION STYLE

APA

Jeng, D. K., Kaczmarek, K. A., Woodworth, A. G., & Balasky, G. (1987). Mechanism of microwave sterilization in the dry state. Applied and Environmental Microbiology, 53(9), 2133–2137. https://doi.org/10.1128/aem.53.9.2133-2137.1987