This study reports a clinical evaluation of AiDEX CGM system featuring a 14-day sensor, real-time glucose monitoring and factory-calibration. A multicenter, prospective, masked clinical study was conducted at with a total of 120 participants. Each participant wore 4 studied sensors and had one in-clinic visit for venous blood reference tests. 40 out of the 120 participants wore additional Abbott Libre sensors and performed at least 7 capillary BG tests daily for additional reference and comparison. Continuous glucose error grid analysis (CG-EGA) showed that AiDEX and Abbott Libre had good agreement with venous blood glucose, with 98.69% and 98.96% accurate readings, respectively. Overall MARD of AiDEX CGM systems was 9.08% when compared to venous blood reference and 10.1% when compared to finger capillary BG reference.

Cite

CITATION STYLE

APA

Ji, L., Guo, L., Zhang, J., Li, Y., & Chen, Z. (2023). Multicenter Evaluation Study Comparing a New Factory-Calibrated Real-Time Continuous Glucose Monitoring System to Existing Flash Glucose Monitoring System. Journal of Diabetes Science and Technology, 17(1), 208–213. https://doi.org/10.1177/19322968211037991