Accurate and precise isotope dilution mass spectrometry method for determining glucose in whole blood

68Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.
Get full text

Abstract

An accurate and precise method to determine glucose concentration in whole blood is presented. The method, based on isotope dilution gas chromatography-mass spectrometry (ID GC-MS), was developed to be used as a Reference Method for determining glucose concentration in capillary or venous whole blood. Blood samples and standards are piperted manually with 'microcap' micropipettes, which makes it possible to collect samples even at the patient's bedside. Glucose is quantified as its aldononitrile pentaacetate. [13C6]Glucose is used as an internal standard. Assay of Seronorm and Pathonorm L and H controls by ID GC-MS gave withinrun CVs of 0.66%, 0.96%, and 0.92%, respectively. For whole blood with glucose concentrations in the low, normal, and high ranges, the within-run CVs were 1.27%, 0.91%, and 0.78%, respectively. The between-run CV for glucose calculated from 36 separate single analyses of Seronorm was 1.44%. In an accuracy assessment test of the HemoCue blood glucose analyzer, 140 capillary blood samples were measured in parallel after split-sampling. For all samples the HemoCue analyzer results had a mean bias of +2.0% compared with the ID GC-MS results.

References Powered by Scopus

The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes Mellitus

0
23925Citations
N/AReaders
Get full text

Consensus statement on selfmonitoring of blood glucose

283Citations
N/AReaders
Get full text

Effectiveness of sodium fluoride as a preservative of glucose in blood

257Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Glucose meters: A review of technical challenges to obtaining accurate results

322Citations
N/AReaders
Get full text

Mass spectrometry in the clinical laboratory

189Citations
N/AReaders
Get full text

A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes

138Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Hannestad, U., & Lundblad, A. (1997). Accurate and precise isotope dilution mass spectrometry method for determining glucose in whole blood. Clinical Chemistry, 43(5), 794–800. https://doi.org/10.1093/clinchem/43.5.794

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 13

57%

Researcher 9

39%

Professor / Associate Prof. 1

4%

Readers' Discipline

Tooltip

Medicine and Dentistry 9

43%

Agricultural and Biological Sciences 5

24%

Chemistry 4

19%

Engineering 3

14%

Save time finding and organizing research with Mendeley

Sign up for free