Flow Cytometry as a Rapid Alternative to Quantify Small Microplastics in Environmental Water Samples

29Citations
Citations of this article
78Readers
Mendeley users who have this article in their library.

Abstract

The most frequently used method to quantify microplastics (MPs) visually by microscope is time consuming and labour intensive, where the method is also hindered by the size limitation at 10 µm or even higher. A method is proposed to perform pre-concentration of MPs by vacuum fil-tration, hydrogen peroxide wet digestion, fluorescent staining and flow cytometric determination to rapidly detect and quantify small MPs sized from 1–50 µm. The method performance was evalu-ated by the spiking of seven different types of polymer, including polystyrene (PS), low-density polyethylene (LDPE), polypropylene (PP), poly(methyl methacrylate) (PMMA), polyvinyl chloride (PVC), polylactic acid (PLA) and acrylonitrile butadiene styrene (ABS) at different levels (400, 4000, 40,000 particles mL−1), with a satisfactory overall % recoveries (101 ± 19.4%) observed, where in general no significant difference between the two methods was observed. Furthermore, a pre-concen-tration process by vacuum filtration was introduced to reduce the matrix effect. After pre-concen-tration, satisfactory % recoveries and accuracy in MP counts resulted from both ultrapure water (94.33 ± 11.16%) and sea water (103.17 ± 9.50%) samples. The validated method using flow cytometry can be used to quantify MPs in environmental water samples that can reduce time and human re-sources.

Cite

CITATION STYLE

APA

Tse, Y. T., Lo, H. S., Chan, S. M. N., & Sze, E. T. P. (2022). Flow Cytometry as a Rapid Alternative to Quantify Small Microplastics in Environmental Water Samples. Water (Switzerland), 14(9). https://doi.org/10.3390/w14091436

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free