70Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The accuracy of the Dinamap 845 automatic blood pressure recorder was assessed by comparing its own indirect determinations of blood pressure with direct intra‐arterial recordings. It was found that in the majority of cases it was capable of producing reliable trend information during anaesthesia. The instrument may not be able to interpret pressure signals from a patient with a severe dysrhythmia. It is probably an unsuitable monitor for use with very rapidly acting drugs such as sodium nitroprusside. Copyright © 1984, Wiley Blackwell. All rights reserved

Author supplied keywords

Cite

CITATION STYLE

APA

HUTTON, P., DYE, J., & PRYS‐ROBERTS, C. (1984). An assessment of the Dinamap 845. Anaesthesia, 39(3), 261–267. https://doi.org/10.1111/j.1365-2044.1984.tb07240.x