We compare remote field eddy current testing and low-frequency eddy current testing from the viewpoint of sensitivity to back-side grooves in a metal plate. We carry out experiments using three metal plates with different depths of artificial grooves. A U-shape magnetic shield gives rise to a remote field eddy current as occurs in pipe inspection. The experiments show that the two methods have almost the same sensitivity to back-side grooves if a relatively low frequency and relatively far coil interval are used, which is also validated by three-dimensional finite-element simulations. We conduct numerical investigations to find the reason for this condition. © 2012 The Japan Institute of Metals.
CITATION STYLE
Wang, J., Yusa, N., Pan, H., Takagi, T., & Hashizume, H. (2013). Evaluation of sensitivity of remote field eddy current testing and low-frequency eddy current testing for inspecting grooves of metal plate. Materials Transactions, 54(1), 90–95. https://doi.org/10.2320/matertrans.M2012323