Alkaline phosphatase based amperometric biosensor immobilized by cysteamine-glutaraldehyde modified self-assembled monolayer

9Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Alkaline phosphatase (ALP) was immobilized with cross-linking agents glutaraldehyde and cysteamine by forming a self-assembled monolayer on a screen printed gold electrode. ALP converts p-nitrophenyl phosphate to p-nitrophenol and phosphate. p-Nitrophenol loses H + ion and turns into the negatively charged compound p-nitrophenolate at medium pH. As a result, the unstable product formed is measured chronoamperometrically at an application potential of + 0.95 V. The biosensor response depends linearly on p-nitrophenyl phosphate concentration between 0.05 0.6 mM with a response time of 40 seconds. Detection limit of the biosensor is 0.033 mM. © 2011 Informa Healthcare USA, Inc.

Cite

CITATION STYLE

APA

Yorganci, E., & Akyilmaz, E. (2011). Alkaline phosphatase based amperometric biosensor immobilized by cysteamine-glutaraldehyde modified self-assembled monolayer. Artificial Cells, Blood Substitutes, and Biotechnology, 39(5), 317–323. https://doi.org/10.3109/10731199.2011.563363