Histological consequences of needle-nerve contact following nerve stimulation in a pig model

26Citations
Citations of this article
33Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background. Nerve stimulation can facilitate correct needle placement in peripheral regional anesthesia. The aim of this study was to determine whether the high threshold current is associated with reduced nerve injury due to fewer needle-nerve contacts compared with low current. Methods. In anaesthetized pigs, thirty-two nerves of the brachial plexus underwent needle placement at low (0.2mA) or high current (1.0mA). The occurrence of needle-nerve contact was recorded. After 48 hours, the nerves were analyzed for occurrence of histological changes. Nerve injury was scored ranging from 0 (no injury) to 4 (severe injury). Results. The frequency of needle-nerve contact was 94 at low compared to 6 at high current. The score was significantly higher at low (median [interquartile range] 2.0 [1.0-2.0]) compared to high current (0.0 [0.0-1.0] P=.001). Conclusions. Inflammatory responses were directly related to needle-nerve contacts. Hence, posttraumatic inflammation may be diminished using higher current for nerve localization. Copyright © 2011 T. Steinfeldt et al.

Cite

CITATION STYLE

APA

Steinfeldt, T., Graf, J., Schneider, J., Nimphius, W., Weihe, E., Borgeat, A., … Wiesmann, T. (2011). Histological consequences of needle-nerve contact following nerve stimulation in a pig model. Anesthesiology Research and Practice, 2011. https://doi.org/10.1155/2011/591851

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free