Reiterative ventricular fibrillation caused by R-on-T during temporary epicardial pacing: a case report

  • Nakamori Y
  • Maeda T
  • Ohnishi Y
N/ACitations
Citations of this article
9Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Epicardial pacemaker wire insertion is standard following cardiothoracic surgery. However, undersensing of pacing wires may cause the R-on-T phenomenon, which induces ventricular fibrillation. We report a case of a male patient with severe mitral regurgitation scheduled for mitral valve replacement who experienced two ventricular fibrillation episodes related to the R-on-T phenomenon caused by undersensing of the epicardial pacing wire. Both undersensing events happened despite an appropriately low sensing threshold. Notably, the stimulated T wave followed the QRS of the premature ventricular contraction (PVC). This case suggests that a PVC's R wave may be undersensed despite a low sensing threshold. This critical complication may have occurred because pacemakers sense R waves using a slew rate, which is the quotient of voltage over time. As a result, pacemakers may undersense wide QRS waves such as PVCs. Avoiding this dangerous phenomenon completely is not possible using epicardial pacemakers; therefore we recommend carefully adapting epicardial pacing especially when PVC waves occur frequently.

Cite

CITATION STYLE

APA

Nakamori, Y., Maeda, T., & Ohnishi, Y. (2016). Reiterative ventricular fibrillation caused by R-on-T during temporary epicardial pacing: a case report. JA Clinical Reports, 2(1). https://doi.org/10.1186/s40981-016-0029-6

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free