Abstract
For in vivo magnetic resonance imaging at high field (≥3 T) it is essential to consider the homogeneity of the active B1 field (B 1+), particularly if surface coils are used for RF transmission. A new method is presented for highly rapid B1+ magnitude mapping. It combines the double angle method with a B1-insensitive magnetization-reset sequence such that the choice of repetition time (TR) is independent of T1 and with a multislice segmented (spiral) acquisition to achieve volumetric coverage with adequate spatial resolution in a few seconds. Phantom experiments confirmed the accuracy of this technique even when TR ≪ T1, with the side effect being lowered SNR. The speed of this method enabled B1+ mapping in the chest and abdomen within a single breath-hold. In human cardiac imaging, the method enabled whole-heart coverage within a single 16-s breath-hold. Results from phantoms and healthy volunteers at 1.5 T and 3 T are presented. © 2006 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Cunningham, C. H., Pauly, J. M., & Nayak, K. S. (2006). Saturated double-angle method for rapid B1+ mapping. Magnetic Resonance in Medicine, 55(6), 1326–1333. https://doi.org/10.1002/mrm.20896