Short report: The CyScope® fluorescence microscope, a reliable tool for tuberculosis diagnosis in resource-limited settings

13Citations
Citations of this article
39Readers
Mendeley users who have this article in their library.

Abstract

Poor laboratory equipment and few human resources have made it difficult to implement microscopic diagnosis of pulmonary tuberculosis (TB) on a large scale basis worldwide. Three hundred sputum samples from patients in Cameroon were studied by using the CyScope®, a new light-emitting, diode-based, fluorescence microscope, to compare auramine-rhodamine fluorescence with the conventional Ziehl-Neelsen staining method. Five fluorescence protocols were tested to reduce manipulation time. Smear positivity for acid-fast bacilli with the Ziehl-Neelsen staining method was 27.7% (83 of 300) compared with 33.3% (100 of 300) with the fluorescent method. Staining time with the modified fluorescence protocol could be reduced from 21 minutes to 10 minutes. This study confirmed that the fluorescence staining method is more sensitive than the Ziehl-Neelsen staining method. It is suggested that the training of laboratory technicians on fluorescence microscopy should be scaled up for increased disease control. Copyright © 2010 by The American Society of Tropical Medicine and Hygiene.

Cite

CITATION STYLE

APA

Lehman, L. G., Yamadji, A. L. N., Sack, F. N., & Bilong, C. F. B. (2010). Short report: The CyScope® fluorescence microscope, a reliable tool for tuberculosis diagnosis in resource-limited settings. American Journal of Tropical Medicine and Hygiene, 83(4), 906–908. https://doi.org/10.4269/ajtmh.2010.10-0153

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free