Robust myoelectric signal detection based on stochastic resonance using multiple-surface-electrode array made of carbon nanotube composite paper

2Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

Abstract

We investigated the robust detection of surface electromyogram (EMG) signals based on the stochastic resonance (SR) phenomenon, in which the response to weak signals is optimized by adding noise, combined with multiple surface electrodes. Flexible carbon nanotube composite paper (CNT-cp) was applied to the surface electrode, which showed good performance that is comparable to that of conventional Ag/AgCl electrodes. The SR-based EMG signal system integrating an 8-Schmitt-trigger network and the multiple-CNT-cp-electrode array successfully detected weak EMG signals even when the subject's body is in the motion, which was difficult to achieve using the conventional technique. The feasibility of the SR-based EMG detection technique was confirmed by demonstrating its applicability to robot hand control.

Cite

CITATION STYLE

APA

Shirata, K., Inden, Y., Kasai, S., Oya, T., Hagiwara, Y., Kaeriyama, S., & Nakamura, H. (2016). Robust myoelectric signal detection based on stochastic resonance using multiple-surface-electrode array made of carbon nanotube composite paper. Japanese Journal of Applied Physics, 55(4). https://doi.org/10.7567/JJAP.55.04EM07