In recent decades, little progress of objective evaluation of pain and noxious stimulation has been achieved under anesthesia. Some researches based on medical signals have failed to provide a general understanding of this problem. This paper presents a feature extraction method for heart rate variability signals, aiming at further improving the evaluation of noxious stimulation. In the process of data processing, the empirical mode decomposition is used to decompose and recombine heart rate variability signals, and the sliding time window approach is used to extract the signal features of noxious stimulation, respectively. The influence of window size on feature extraction is studied by changing the window size. By comparing the results, the feature extraction in the process of data processing is valuable, and the selection of window size has a significant impact. With the increase of selected window sizes, we can get better detection results. But for the best choice of window size, to ensure the accuracy of the results and to make it easy to use, then, we need to get just a suitable window size.
CITATION STYLE
Yin, Q., Shen, D., & Ding, Q. (2021). Influence of Sliding Time Window Size Selection Based on Heart Rate Variability Signal Analysis on Intelligent Monitoring of Noxious Stimulation under Anesthesia. Neural Plasticity, 2021. https://doi.org/10.1155/2021/6675052