Abstract
A sensitive method based on a geometry-independent neutral desorption (GIND) in combination with extractive electrospray ionization mass spectrometry (EESI-MS) has been developed for fast detection of illicit additives such as sulfonamides and hormones in highly viscous cosmetic products. The method gave a low limit of detection (LOD) (in the range of 0.001-1 ng/g), acceptable relative standard deviation (RSD=6.8-11.4%) and reasonable recovery (87-116%) for direct measuring of nine types of hormones and sulfonamides in the cosmetic products. The average measurement time for two types of samples was less than 1 min. Trace amounts of analytes in commercial cosmetic products have been quantitatively detected, without any sample pretreatment. The experimental results showed that non-volatile illicit additives such as sulfonamides and hormones could be sensitively liberated using the GIND device for quantitative detection from the highly viscous cosmetic products, demonstrating that GIND-EESI-MS is a promising tool for high throughput, sensitive and quantitative analysis of highly complex viscous samples. Copyright © 2011 John Wiley & Sons, Ltd.
Cite
CITATION STYLE
Liu, Y., Zhang, X., Ouyang, Y., Hu, Z., Ma, L., Zhang, J., … Chen, H. (2011). Trace detection of hormones and sulfonamides in viscous cosmetic products by neutral desorption extractive electrospray ionization tandem mass spectrometry. Journal of Mass Spectrometry : JMS, 46(8), 794–803. https://doi.org/10.1002/jms.1944