Measurement of lung function using Electrical Impedance Tomography (EIT) during mechanical ventilation

4Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The consistency of regional lung density measurements as estimated by Electrical Impedance Tomography (EIT), in eleven patients supported by a mechanical ventilator, was validated to verify the feasibility of its use in intensive care medicine. There were significant differences in regional lung densities between the normal lung and diseased lungs associated with pneumonia, atelectasis and pleural effusion (Steel-Dwass test, p < 0.05). Temporal changes in regional lung density of patients with atelectasis were observed to be in good agreement with the results of clinical diagnosis. These results indicate that it is feasible to obtain a quantitative value for regional lung density using EIT. © 2010 IOP Publishing Ltd.

Cite

CITATION STYLE

APA

Nebuya, S., Koike, T., Imai, H., Noshiro, M., Brown, B. H., & Soma, K. (2010). Measurement of lung function using Electrical Impedance Tomography (EIT) during mechanical ventilation. In Journal of Physics: Conference Series (Vol. 224). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/224/1/012029