Ultrafiltration discrepancies in recovery of myoglobin from urine

12Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We investigated the efficiency accuracy, and reliability of the ultrafiltration/dipstick methodology commonly used to diagnose myoglobinuria. Twenty-five myoglobin-containing urine specimens were filtered by centrifugation for 15 min at 1500g through a Centricon-30 membrane filter. Both the original specimen and filtrate were assayed for myoglobin. The amount of myoglobin recovered subsequent to filtration varied from <60 000 μg/L. Fourteen of 18 such filtrates had concentrations <350 μg/L, a concentration below which a negative result would be obtained by using conventional dipstick methods. Thus, the use of this procedure has the potential to misdiagnose patients with myoglobin concentrations associated with increased risk of subsequent renal dysfunction, in particular when urine myoglobin concentrations are <60 000 μg/L.