Over one-year long-term laboratory tests of ph electrodes in terms of industrial applications checking stabilities of their parameters and their influence on uncertainties of measurements

10Citations
Citations of this article
42Readers
Mendeley users who have this article in their library.

Abstract

Glass pH electrodes are still successfully applied in the chemical and environmental industry. During their long-term use, periodic calibration is required to maintain the required accuracy of measurements because the parameters of the electrodes change over time. This work presents an aging of 11 pH electrodes within approximately 600 days in tap water. During this period, potentials of all electrodes in five buffer solutions were measured 44 times. This allowed determining the aging models of the electrodes. Models in other mediums might be different. Changes in slope, standard potential, hysteresis, and linearity of the characteristics were the objects of observations. A method for predicting electrode parameters is proposed. Furthermore, the procedure for estimating the uncertainty of pH measurements considering the aging of the electrodes is described. As a result of this work, a model of the aging process of pH electrodes in tap water can be constructed and subsequently, the measurement accuracy in the periods between calibrations can be improved.

Cite

CITATION STYLE

APA

Wiora, A., & Wiora, J. (2018). Over one-year long-term laboratory tests of ph electrodes in terms of industrial applications checking stabilities of their parameters and their influence on uncertainties of measurements. Sensors, 18(12). https://doi.org/10.3390/S18124102

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free